10
P1.0(rEc) Recording
P1.1(int) Interval
P1.2(rES)
Reservation start
1,rES ON or OFF select
1A~1E, start date to minute
P2.0(AL) Temperature alarm
P2.1,CH1 to Ch4 select
P2.2, High or Low select
P2.3, ON or OFF select
P2.4, alarm threshold
P3.0(oFSt) offset
P3.1,CH1 to Ch4 select
P3.2, ON or OFF select
P3.3, offset value
P4.0(dAt) Date, real time clock
P4.1,time format
P4.2~4.6, rtc input
P5.0(UNit) Unit
P5.1,C/F select
Sampling rate starts from 1 second to 90 minutes
First, choose to turn ON or OFF the reservation function.
While it is ON, setup the start Date & Time
WARNING: Reservation start time CAN NOT be
earlier than NOW
First, choose the channel and decide high or low alarm
threshold.
Then, each high or low alarm can be enabled or disable
by choosing ON or OFF.
The previous alarm threshold value is stored in meter,
the last step is to setup the threshold. The adjustable
range is -200C to 1370C
First, choose the channel. Then, each set value can be
enabled or disable by choosing ON or OFF.
The previous offset value is stored in meter, you may
use P3.3 to change the value again. The adjustable
range is +/-12C
First, choose the time format as 12H or 24H.
Then, setup the time from year, month, day, hour to
minute
Select the temperature unit as C or F. Default is C.
Program