background image

- 8 -

Measuring the tested object that is already removed from the in-line circuit can 

the in-line circuit can help to obtain a more accurate reading.

D.Testing Diodes

 (See Figure 6)

 red      black

( Figure 6)

Warning

To avoid damages to the Meter or to the devices under

test, disconnect circuit power and discharge all the

high-voltage capacitors before testing diodes.

Use the diode test to check diodes, transistors, and other

semiconductor devices. The diode test sends a current

 through the semicondutor junction, then measure the

voltage drop across the junction. A good silicon junction

drops between 0.5V and 0.8V.

To test the diode out of a circuit, connect the Meter as follows:

 1.  Insert the red test lead into the

V

Ω

terminal and

      the black test lead into the 

COM

 terminal.

 2.  Set the rotary switch to
 3. For forward voltage drop readings on any semiconductor component, place the red

 test lead on the component’s anode and  the black test lead on the component’s cathode.

Summary of Contents for MSR-C600

Page 1: ...Digital Clamp Multimeter Model No MSR C600 ...

Page 2: ...he package case and take out the Meter Check the following items carefully for any missing or damaged part This Meter complies with the standard IEC61010 Pollution Degree 2 Overvoltage Category CAT II 600V CAT III 300V and Double Insulation CAT II Local level appliance PORTABLE EQUIPMENT etc with smaller transient overvoltages than CAT III CAT III Distribution level fixed installation with smaller...

Page 3: ... 2 Warning range ...

Page 4: ...t alter the internal circuit of the Meter randomly used AC Alternating Current DC Direct Current AC or DC Grounding Double Insulated Continuity Test Diode Fuse Warning Refer to the Operating Manual Conforms to Standards of European Union Deficiency of Built In Battery Danger of High Voltage ...

Page 5: ...ow table indicated for information about the rotary switch positions Functional Buttons Below table indicated for information about the functional button operations Rotary switch positions Function DC voltage measurement AC voltage measurement Resistance measurement Continuity diode test Ω Power is turned off OFF AC current measurement from 0 001A to 400A 2 20A A 200 400A A Press HOLD to enter and...

Page 6: ...r AC voltage or current 2 DC Indicator for DC voltage MAX Maximum reading displayed Ω Ohm The unit of resistance kΩ kilohm 1 x 103 or 1000 ohms MΩ Megaohm 1 x 106 or 1 000 000 ohms Ω kΩ MΩ 9 10 A Amperes amps The unit of current mV V Volts The unit of voltage mV Millivolt 1x10 3 or 0 001 volts 11 12 Indicates negative reading 13 OL The input value is too large for the selected range Warning To avo...

Page 7: ...ject being measured The measured value shows on the display red black Figure3 Warning To avoid harms to you or damages to the Meter from eletric shock do not attempt to measure voltages higher than 600V AC DC although readings may be obtained VΩ V B Measuring AC Voltage See Figure 4 red black Figure 4 Warning The AC Voltage ranges are 2 000V 20 00V 200 0V and 600V To avoid harms to you or damages ...

Page 8: ... the Meter as follows 1 Insert the red test lead into the VΩ terminal and the black test lead into the COM terminal 2 Set the rotary switch to V 3 Connect the test leads across with the object being measured The measured value shows on the display ...

Page 9: ...o check diodes transistors and other semiconductor devices The diode test sends a current through the semicondutor junction then measure the voltage drop across the junction A good silicon junction drops between 0 5V and 0 8V To test the diode out of a circuit connect the Meter as follows 1 Insert the red test lead into the VΩ terminal and the black test lead into the COM terminal 2 Set the rotary...

Page 10: ... 9 Measuring the tested object that is already removed from the in line circuit can help to obtain a more accurate reading See Figure 7 ...

Page 11: ...witch or press any button for around 15 minutes The Meter can be activated by turning the rotary switch or pressing any button with the following conditions The Hold function will be cancelled if the Meter is activated by pressing the HOLD button To disable the Sleep Mode function press and hold HOLD button while turning on the Meter Sleep Mode ...

Page 12: ... 11 Low Battery Indication Display 6 1 Specifications Opening Tested H 2 ...

Page 13: ... 12 3 Display sinewave RMS AVG response ...

Page 14: ... 13 Maintenance This section provides basic maintenance information including battery replacement instruction sinewave RMS AVG response RMS value ...

Page 15: ...ry 6 Rejoin the case bottom and the battery compartment and reinstall the screw 14 Figure 9 Screw Warning A General Service B Replacing the Battery See Figure 9 Do not attempt to repair or service your Meter unless you are qualified to do so and have the relevant calibration performance test and service information To avoid electrical shock or damage to the Meter do not get water inside the case P...