GEN04
User Manual
12
Error Code 1: Generator
Error Code 1 indicates either there is a functional problem with the
generator or the front panel button(s) were activated during power-up
sequence.
Cycle the power OFF then ON. If error persists, power off system and
contact service.
Error Code 2: Generator
Error Code 2 indicates that the generator is overheating.
Temperature
1
Power off system. Remove any obstructions blocking the air
vents on the generator’s bottom and back panels. If there is no
apparent obstruction or external heat source, contact service.
2
Power on the system and wait for up to 30 minutes for generator
error to clear.
3
If error code persists, contact service.
Error Code 3: Hand Piece
Error Code 3 indicates a problem with the hand piece.
1
Confirm that the hand piece connector is fully inserted and
properly oriented – white dot on handpiece is aligned with white
dot on front panel. If the error code does not clear within three
seconds after the hand piece is properly connected, press TEST.
2
The instrument may not be tightened properly or tissue may
have collected in the distal end of the instrument shaft. Tighten
instrument using blade wrench and carefully remove tissue from
distal end of instrument sheath. Press STANDBY to clear error
code and return to Ready mode. Activate system. If the pre-run
test is running, ensure instrument is in air. If using shears, ensure
jaws are open and not in contact with any objects during pre-
run test.
Note: Inspect the blade wrench hub for cracks or wear before use. If
damage is seen, replace the blade wrench. Before use after
autoclaving, cool the blade wrench at room temperature for at least 45
minutes or soak it in room temperature sterile water for 5 minutes.
3
If the error persists, install a test tip to isolate the problem. Press
TEST button.
-
If system indicates a hand piece error, hand piece is bad or
test tip is bad. Replace hand piece or install a new test tip
and press TEST.
-
If no error occurs with test tip attached, replace instrument.
4
Press STANDBY to return to Ready mode. Activate system.
Error Code 4: Hand Piece
Error Code 4 indicates that the hand piece has exceeded its
Temperature
specified operating temperature. For immediate recovery, use another
hand piece; or, follow the steps below to determine the cause of the
error condition and alternate recovery methods.
Summary of Contents for ULTRACISION HARMONIC SCALPEL Generator 300
Page 2: ...GEN04 ...
Page 4: ...GEN04 ...
Page 6: ...User Manual 2 GEN04 ...
Page 20: ...User Manual 16 GEN04 ...
Page 22: ...User Manual 18 GEN04 ...
Page 24: ...User Manual 20 GEN04 ...
Page 30: ...User Manual 26 GEN04 ...
Page 32: ...Manufacturer Authorized Representative in the European Community User Manual 28 GEN04 ...
Page 33: ...Chapter 10 Symbols 29 GEN04 ...
Page 34: ...Manuel d utilisation 30 GEN04 ...