6
For example: Set over voltage protection value to 33.0V.
Press
[OVP] [3] [3] [.] [0] [Enter]
.
3.5.2 OVP Status Setting
Press
[Menu]
button
until “OVP Status” is shown on the LCD
Press buttons
[
]
or
[
]
to turn on or off the OVP function.
3.6 Over Current Protection Setting
The Over Current Protection (OCP) function protects the power supply and DUT from damage caused by current
over the setup current. Before operation, set the OCP function ON and set the OCP value. When the output current
exceeds this value, the output will be shut down and
“
OP
”
indicator will light on. With OCP function ON, the
output current is restricted within the OCP range.
3.6.1 OCP Value Setting
Press
[OCP]
button
until “OCP Set” is shown on the LCD.
Setting method 1: Press
[numeric buttons 0-9]
[Decimal Point] [Enter]
to complete OCP setting.
Setting method 2: Press buttons
[
]
or
[
]
to move the cursor left or right. Rotate the rotary knob left or right to
increase or decrease the digit on cursor.
For example: Set over current protection value to 3.30A.
Press
[OCP] [3] [.] [3] [0] [Enter]
.
3.6.1 OCP Status Setting
Press
[Menu]
button
until “O
C
P Status” is shown on the LCD
.
Press buttons
[
]
or
[
]
to turn on or off the OVP function.
3.7 Output Delay Time Setting
Press
[Menu]
button until
“Delay” is shown on the LCD.