11
3.9.3 Cycle Times Setting
Press
[Menu]
button
until “Auto Cycle” is shown on the LCD.
Setting method 1: Press
[numeric buttons 0-9] [Enter]
to set cycle times from number 0 to 99999. Number 0
represents infinite loop.
Setting method 2: Press buttons
[
]
or
[
]
to move the cursor left or right. Rotate the rotary knob left or right to
increase or decrease the digit on cursor.
For example: Set the cycle times to 99.
Press
[9] [9] [Enter]
.
3.9.4 Enter/Exit Auto Running Operation
Press
[ListRun]
to enter or exit the auto running operation mode. If the Auto key light turns on, it means the
instrument has entered the auto running operation.
Under this mode,
“
LeftTime
”
shown on the LCD represents the left time of running settings on the present
memory address;
“
LeftCycle
”
represents left executed cycles;
“
No.
”
means the memory address.
Press [On/Off] mode to start output.
3.10 Output On/Off
The Output
[On/Off]
button on front panel is used to turn on or off the output.
3.11 Panel Lock
The
[Lock]
button on front panel is used to lock the front panel from misinput. Press
[Lock]
button, then the
button backlight lights on, meaning the front panel is locked. All the other buttons and rotary knob are disabled
except
[Lock]
button and
[On/Off]
button.
3.12 Rotary Knob
Press the rotary knob to switch the cursor position.
3.13 Protection Function
This power supply is built with Over Voltage Protection (OVP), Over Current Protection (OCP), Over Power
Protection (OPP) and Over Temperature Protection (OTP).
OVP and OCP level can be set via the [OVP-set] and [OCP-set] buttons on the front panel respectively.