10
|
Maintenance
Battery Maintenance
Under normal operating conditions, recharging the battery is the only
HI-3702 maintenance performed by the user.
Maintaining the HI-3702 requires that the nickel-cadmium (NiCd) battery be
properly charged and the battery condition monitored. The normal operational life
of the battery (approximately 1000 charge/discharge cycles) is such that by the
time the battery needs replacement; the HI-3702 will need recalibration; these
tasks are performed by ETS-Lindgren. For information on calibration services,
see
Annual Calibration
on page 10.
For the steps to charge the battery, see page 20.
In case of damage to the battery that causes premature failure, the HI-3702 must
be sent to ETS-Lindgren for battery replacement; replacement in the field is not
possible. See
Service Procedures
on page 11 for information on returning the
HI-3702 to ETS-Lindgren for service.
Annual Calibration
See the
Product Information Bulletin
included with your shipment for information
on ETS-Lindgren calibration services.
Maintenance of Fiber Optics
Fiber optic connectors and cables can be damaged from airborne particles,
humidity and moisture, oils from the human body, and debris from the connectors
they plug into. Always handle connectors and cables with care, using the
following guidelines.
Summary of Contents for HI-3702
Page 1: ...Model HI 3702 Clamp On Induced Current Meter User Manual ...
Page 15: ...Ranges and Specifications 15 HI 3702 Typical Frequency Response ...
Page 16: ...16 Ranges and Specifications This page intentionally left blank ...
Page 22: ...22 Operation This page intentionally left blank ...
Page 24: ...24 Functional Theory of Operation This page intentionally left blank ...
Page 26: ...26 Warranty This page intentionally left blank ...
Page 32: ...32 Series H 491198 36 Battery Charger This page intentionally left blank ...
Page 34: ...34 EC Declaration of Conformity SERIES H 491198 36 BATTERY CHARGER ...