5.  Plug in the humidifier. 
 

A.  If an ETS Automatic Humidity Controller (Model 5100 or Model 5200) 

is being used to control the humidity, 

plug the humidifier into the AC 

outlet on the rear of the unit labeled INCREASE.

 

 
B.  If no controller is being used, then plug the unit into a standard AC 

outlet of appropriate voltage. 

 

NOTE 

230 VAC Model 5462 Systems :

 The 230 VAC version of the 

Model 5462 must be operated with the provided 230 VAC to 115 
VAC step-down transformer 

ONLY!!!

  Plug the transformer into a 

230 VAC outlet (or 230 VAC humidity controller output), then plug 
the humidifier into the 115 VAC side of the transformer.  NEVER 
plug the humidifier directly into a 230 VAC outlet, the humidifier is 
meant for operation at 115 VAC only. 

 

The Model 5472 is available as 115 VAC or 230 VAC, 50/60 Hz  The 
voltage must be specified at the time of purchase

.   

 
   The circulating fan is supplied at the appropriate voltage, 115 VAC  
   or 230 VAC. 

 

 

4.0 MODEL 5462 

 

 

The Model 5462 Ultrasonic Humidifier is a direct replacement for the ETS 
Model 562 or 5612 Ultrasonic Humidifier.   

 

4.1     SET-UP 
 

1.  Fill the water tank.  

Distilled or Deionized water ONLY!!

 

 

Remove the water tank from the humidifier unit and inspect it for small 
cracks or any other damage that may have occurred during shipping (a 
small crack will allow air to enter the tank, which can cause the water to 
overflow the basin and possibly damage the unit).  After inspection, fill the 
tank ¾ full with distilled or deionized water ONLY (user provided). 
 

        NOTE:  Using tap water will destroy the ultrasonic transducer and the  

associated electronics.  Tap water will also cause a white dust  
to form on all surfaces (including the humidity sensor, which 
will also be destroyed). 

 
 

Summary of Contents for 5462

Page 1: ...Ultrasonic Humidification System Model 5462 Model 5472 OPERATING MANUAL 6 30 06 www electrotechsystems com...

Page 2: ...l 5462 03 4 2 Model 5462 Operation 04 5 0 Model 5472 06 5 1 Model 5472 Set Up 07 5 11 Using the Water Tank 07 5 12 Using the Tap Water System 09 5 2 Operation 10 5 2 1 Features 10 5 22 Operating Preca...

Page 3: ...late producing a fine mist An air moving device forces the mist out of the humidifier through a 1 plastic tube and into the enclosure A 4 110 cfm fan is provided it should be run continuously to circu...

Page 4: ...e the chamber If the humidifier is being used with an ETS Chamber the barb is already installed 2 Place the humidifier a minimum of 12 away from the chamber 3 Cut the 1 I D tubing to the appropriate l...

Page 5: ...is available as 115 VAC or 230 VAC 50 60 Hz The voltage must be specified at the time of purchase The circulating fan is supplied at the appropriate voltage 115 VAC or 230 VAC 4 0 MODEL 5462 The Mode...

Page 6: ...ist 4 Rotate the MIST INTENSITY on the POWER switch This control determines how fast the water is converted from a liquid to a mist For most applications set the control at mid point II position Rotat...

Page 7: ...mpartment 3 Never place the humidifier inside the chamber The fans and electronics will become exposed to the wet air and will quickly fail Always operate the humidifier outside the chamber and attach...

Page 8: ...Ultrasonic Humidifiers The Model 5472 has many features not found in the Model 5462 Figure 5 0 1 5 Gallon Water Tank for long term applications Deionizer Filtering Column 10 ft OD Polytubing Quick Co...

Page 9: ...deionizing resin is depleted it will change from dark to light in color When it is light it must be replaced 5 11 Using the Water Tank 1 Attach the water de ionizing column to a faucet The faucet shou...

Page 10: ...the humidifier 5 Attach the provided plastic faucet to the water tank The faucet has a O D tubing quick connect fitting attached to the outlet 6 Make sure the faucet is closed then push the O D tubing...

Page 11: ...urified water Fast flow rate poorly purified water 3 Turn on the water Set the flow rate to the pre determined amount DO NOT OPERATE THE HUMIDIFIER WITH THE FAUCET 100 OPEN High flow rates will cause...

Page 12: ...er in the Model 5472 3 The Model 5472 can be modified to use compressed air or an inert gas as the propulsion source instead of the air pump The 5472 can also be modified for to be a close loop system...

Page 13: ...y 6 0 MAINTENANCE CLEANING Model 5462 5472 1 Always unplug the unit before cleaning 2 Empty the unit of all water Disconnect or remove water tank or tap water source 3 Clean the surface of the transdu...

Page 14: ...l parts of the humidifier when it is to be stored or taken out of service for any period longer than one week Do not seal the water tank in storage Leave the top off to allow the air to completely dry...

Page 15: ...anty becomes null and void should the equipment or any part thereof be abused or modified by the customer of if used in any application other than that for which it was intended This warranty to repla...

Page 16: ...TANTIALLY INCREASES THE CONTENT S SUSCEPTIBILITY TO SHOCK DAMAGE DO NOT PLACE INSTRUMENTS OR ACCESSORIES INSIDE OTHER INSTRUMENTS OR CHAMBERS ELECTRO TECH SYSTEMS INC WILL NOT ASSUME RESPONSIBILITY FO...