Model 5503 Compact Chamber
–
Operating Manual
D01307 Revision B - 2021-03-16 - Page
10
of
20
Step 9
–
Humidifier
– Connect to chamber.
•
Plug the AC cord into
the “Humidity Increase”
socket of the controller.
•
Insert 1” mist tubing from top of humidifier to
chamber 1” fitting.
•
Set a set point by holding down
the “
*
” key on
the controller and using the
increase//decrease
(▲//▼) buttons to enter
the desired humidity.
•
Switch the front panel Humidify switch to ON.
The controller will request humidification as
needed.
Consult the Operating Manual for the 5482
Humidification System for more details.
Step 10
–
Dehumidifier Model 5461 Setup
• Desiccator column should be filled with
Molecular Sieve desiccant, white with blue
indication pellets.
• Place the desiccator column, lid down, onto
the 4 mounting studs of the base. The side
marked “IN” should face forward. Marking is on
the bottom of the column.