Model 5503 Compact Chamber
–
Operating Manual
D01307 Revision B - 2021-03-16 - Page
18
of
20
VII. Maintenance and Calibration
Calibration
For consistency of performance Annual manufacturer conducted calibration is required.
ETS recommends the Model 5503 system be calibrated annually. Only the controller and sensor need
to be returned for calibration. Do NOT return the chamber.
To return equipment to ETS for calibration or repair it is first necessary to obtain a RMA number,
please call 215-887-2196 or email
Preventative Maintenance
Humidification System
The humidification system utilizes an ultrasonic transducer that has an estimated lifespan of 3000
hours after which it may need to be replaced. The ultrasonic transducer is not user replaceable.
Please contact ETS for assistance with repair/replacement. To maximize the life of the transducer,
utilize an appropriate water supply (deionized water or distilled water) and perform regular
preventative maintenance after every
300 hours
of operational use.
Humidification System Preventive Maintenance
Step 1
– Turn off power
Unplug the AC line cord from the AC outlet of
controller to remove power from the humidifier.
Step 2
– Prepare and drain the basin
The water will need to be drained out of the
basin first. First, place the unit above a small
container at the end of the drain valve to
collect drained water. Turn the drain valve to
ON position [tap turned parallel to the valve
body]. Wait for the water to stop flowing
before proceeding.
Step 3
– Turn on power
Empty the small container, and place back
until the drain; leave the drain valve open.
Plug the AC line cord in a wall AC outlet to
apply constant power to the humidifier.
POWER OFF