Model 5506 Mid-Size Chamber
–
Operating Manual
D01332 Revision B - 2021-03-16 - Page
9
of
20
Step 6 -
Connect Fan and Outlet
•
Adjacent to the Heater power connection are
the power input plugs for the Fan and the
Outlet. These are labeled on the left side of
the heater panel.
•
Connect power cables to these inputs and
plug them directly into AC outlets.
Step 7
–
Humidifier Model 5482
– Set up Water
Source
•
With tank dispensing valve OFF (right), fill the
water tank with 4 gallons of de-ionized water.
•
Position the tank above the base unit and be
sure the base unit is on a level surface.
•
Connect the water feeding tube from the 5-
gallon tank to the humidifier as shown, inserting
tubing firmly into the connectors.
•
Turn the tank dispensing valve to ON (left)
•
Loosen the small white cap on the top of the
tank to allow air to flow and pressure to relieve.
Step 8 -
Humidifier
– Base unit startup run
•
Insert 1” diameter mist tube in the top fitting on
the base unit.
•
Plug the three-prong AC line cord into an outlet
for a startup test.
•
LED will turn
RED
as basin fills, 1 to 2 mins.
•
LED will turn
GREEN
and mist will begin.
•
After 3 seconds, mist output will increase to
maximum level.
•
Disconnect the AC cord.