U2836 Operation Manual Chapter 3 Introduction to [MEAS DISP]
28
On
<LIMIT TABLE >
page, no low / high limit of the secondary parameters
has been set.
On
< LIMIT TABLE >
page, the low/high limit of secondary parameters has
been set but
AUX BIN
function is set as OFF.
In this case, only those components whose secondary parameters are qualified
can perform primary parameter sorting according to sorting limits. If the
secondary parameters are unqualified and the corresponding primary parameters
are within limit ranges, those components will be sorted into BIN OUT.
On <
LIMIT TABLE
> page, the low/high limit of the secondary parameters
has been set and the
AUX BIN
function is set as ON.
If the primary parameter is out of the limit range, it is sorted into BIN OUT. If the
primary parameter of DUT is within the limit range but its secondary parameter is
out of the limit range, the DUT will be sorted into the
AUX BIN
bin.
———————————————————————————————————————
Note: When the secondary parameter only has low limit and the auxiliary bin is
set as ON, if the primary parameter of DUT is within the limit range and the
secondary parameter is smaller than or equal to its low limit, the DUT will be
sorted into the auxiliary bin. When the secondary parameter only has high limit
and the auxiliary bin is set as ON, if the primary parameter of DUT is within the
limit range and the secondary parameter is larger than or equal to its high limit,
the DUT will be sorted into the auxiliary bin.
———————————————————————————————————————
Operation steps for setting the auxiliary bin function ON/OFF
1)
Move the cursor to
AUX
, the following soft keys will be displayed.
ON
OFF
2)
Use above soft keys to set the auxiliary function as ON or OFF.
3.5.6
HIGH/LOW
U2836 can set bin limits of 3 primary parameters and one secondary parameter.
The test results can be sorted into 4 bins at most (BIN 1 to BIN 9 and BIN OUT).
The high/low limits of primary parameters can be set in high limit and low limit of
bins from BIN 1 to BIN 3. The limit of the secondary parameter can be set in
HIGH
and
LOW
of 2
nd
.
Operation steps for setting high/low limit
Execute the following steps to set sorting limits.
1)
Set
FUNC
and
NOM
in the compare function menu and the limit
MODE
of the
primary parameter.
2)
Move the cursor to Low limit of BIN 1. If you select percent tolerance mode,
the following operation steps should be from step 3 to step 6; if you select
absolute tolerance mode and sequential mode, the following operation steps
should be from step 7 to step 11.