Replacing the fuse
To change the fuse, following the procedure below:
1.
Unplug the system from power and place microscope flat, with base of microscope toward
you.
2.
Find the fuse cover that will appear as a round protrusion with a slot.
3.
Use a small flathead screwdriver or other flat object (coin, etc) to gently push the fuse
cover in and turn the cover counter clockwise. You need to turn the cover about 3/4 of a
turn.
4.
The fuse cover will pop out with the fuse attached.
5.
Remove the fuse from the cover and examine the fuse. If the thin piece of metal going
from one end of the fuse to the other has a gap, then the fuse is bad.
6.
If the fuse is bad, install a replacement fuse in the cover.
7.
Gently push the fuse cover with the new fuse back into the sub-woofer until it is flush with
the unit. Turn the cover clockwise about 3/4 to secure the cover back into the unit.
Note:
Fuse may blow in order to protect internal damage to the microscope. And in most cases, replacing the fuse with the correct
voltage will resolve the issue. However, should you encounter a blown fuse frequently, please contact your distributor for further
assistance.
fuse specification: 250V, 150 mA