12.2 Impulsed dosing
In order to set the flowmeters it is necessary to control that the chemical products dosing system is completely filled.
12.2.1 Calibration
Enter the menu: SETTINGS
→
→
→
→
FLOWMETERS
→
→
→
→
Insert 2
nd
level password
→
→
→
→
CALIBRATION
Select the flowmeter that you want to calibrate by using P1 and P2 buttons.
ATTENTION: If it is not present the flowmeter the system displays the message “FORBIDDEN”.
Insert the suction lance of the chemical to be calibrated into a ml-graduated cylinder and fill it with the chemical up to
250 ml.
Chemical product
After selecting the flowmeter to calibrate (see the input output section 11.6) push START to start the procedure and
CONFIRM it.
Press START when the level on the ml-graduated cylinder have reached the displayed quantity (1).
If you want to shut off the procedure press STOP.
Chemical product
1
250 ml
150 ml