REV.0.02_COD.500201_A4
PAG. 67
CLEANING OF DRYING HEPA FILTER (IF PRESENT)
M5
Worker: Is
Frequency of Intervention: 300 hours
METHOD OF INTERVENTION: clean (or replace) the drying system filter as described below:
•
Remove the two screws from the drying system filter protective front panel and remove it from the machine.
•
Extract the pre-filter and than the HEPA filter: replace it with another filter of the same type.
•
Carefully put the pre-filter back in place. Use the screws to fasten the previously removed protective front panel.