100 MM GRAPHICS RECORDER: USER GUIDE
User Guide
Page 11
HA027271
Issue 12 Mar 04
2.2.2 Supply voltage wiring
WARNING
DC supply voltages must never be applied to recorders fitted with isolated transmitter power supplies.
Note: The minimum recommended wire size is 16/0.2 (0.5mm
2
).
LINE SUPPLY
The supply voltage to the recorder is terminated using an IEC socket which is connected to the mating plug at the rear
of the recorder. The recorder is suitable for use with all ac voltages between 85 and 265 V RMS (47 to 63 Hz), and
requires 60 VA max. power. For recorders without transmitter power supplies, supply voltages of between 110V dc
and 370V dc may also be used.
Figure 2.3a Disk access
Figure 2.3b Disk eject
LOW VOLTAGE SUPPLY OPTION
Not suitable for recorders fitted with the isolated transmitter power supply
option.
The low voltage supply option is terminated at a three-pin connector (plug
mounted on recorder - socket on supply cable) as shown in figure 2.2.2.
The option allows the use of ac or dc supplies with the following charac-
teristics:
AC:
20 to 42V RMS (45 to 400 Hz)
DC:
20 to 54V (See warning above)
Power: 60VA max.
2.3 DISK INSERTION AND REMOVAL
The disk slot is located below the recorder screen, and is protected by a flap as shown in figure 2.1, above.
In order to access the disk slot, the bottom of the central part of the flap is lifted (figure 2.3a) and then used as a han-
dle to pull the main flap open.
If a disk is already fitted, it is removed by pressing on the eject button (figure 2.3b).
Note: Before disk removal, archiving should be suspended (
) (wait for the green LED on the disk
drive to be extinguished), otherwise data might be lost.
+V or ac
Earth
0V or ac
Figure 2.2.2 Low voltage supply pinout