Immediately after attaching the
infusion set: Remove and dispose
of the vial
Grip the vial by the adapter and
remove both vial and adapter from
the reservoir by turning them to the
left (counterclockwise). Dispose of
these components with the household
waste in accordance with
Section
9.5
.
Connect infusion set
Check to ensure that the threaded
connection on the reservoir is still
clean. Remove a compatible infusion
set from the packaging and connect it
to the reservoir of your pump.
Tighten the threaded connection
securely to ensure that the infusion set
does not loosen during drug delivery.
CONNECTING,
PREPARING,
AND PRIMING THE
INFUSION SET
3.2
CONNECTING
THE INFUSION
SET
Confirm the procedure with
.
If the infusion set connection is not
adequately tightened, it could start
leaking, leading to an incorrect
dose or no delivery of the drug.
Overtightening the threaded
connection can damage the
connection point.
If the port on the reservoir is soiled,
replace the soiled reservoir with a
new one.
Only use infusion sets with the
following characteristics:
Needle diameter between 28 and
31 gauge
To proceed immediately with the
connection of an infusion set, remove
the vial and press
Yes
.
If you wish to connect the infusion set
at a later time, then leave the vial on
the pump until that time and press
No
. This will take you to the main
screen, where you can then select
"Change infusion set" in the menu in
Section
4.3
to proceed.
After removing the vial, the port
for the infusion set is readily
accessible. Avoid touching the
port or any contact between the
port and other objects to ensure
that the port remains sterile.