2
Internal connection External wired connection
You can choose either the internal or external wired connection, both of which types
cannot be in existence at the same time.
Note:
After removing batteries, wait for 5 seconds to refit batteries.
Operation
1. With the tamper switch not being pressed, the unit enters test mode, which
allows the user to make a test. Detach the magnet from the Detector, the red
indicator LED on the Detector will illuminate and the load or the lamp plugged
into the On/Off Module AN135, Lamp Module AD130 will turn ON. It implies
that the unit is working properly.
Close the door/window by replacing the magnet to the Detector will turn OFF
the load or the lamp plugged into the On/Off Module AN135, Lamp Module
AD130 and the red indicator LED on the connected load or lamp will be off too.
2. When the tamper switch is pressed, the unit enters normal mode. Open the
door/window to detach the magnet from the Detector will turn ON the load or the
lamp plugged into the On/Off Module AN135, Lamp Module AD130.
Close the door/window by replacing the magnet to the Detector will turn OFF
the load or the lamp plugged into the On/Off Module AN135, Lamp Module
AD130.
In normal mode with the tamper switch being pressed, the red indicator LED on
the Detector will not illuminate to conserve battery life when the detector is
triggered, (unless the battery is low).
3. By pressing the tamper switch for more than 5 seconds, and then release it.
The Detector will send an alarm command
(ALARM_REPORT, Alarm Type ==
0x01, Alarm Level == 0x11)
to the module, of which load or the lamp plugged
into the On/Off Module AN135 or Lamp Module AD130 will flash for 10 seconds.
Advanced Operation
The following information is for someone that has some experience setting up a
Z-Wave system or someone that has computer software running a Z-Wave
controller.
Enabling/Disabling Power Saving Function (for testing)
When the magnet is parted or closed from the Detector for 10 seconds, the SM103
will enter the power saving mode. It can be disabled or enabled power saving
function by setting Configuration Parameter # 3.
This parameter can be configured with the value of 0 through 127, where 0 means
power saving being enabled and others mean power saving being disabled.
PS : As long as the batteries have been refitted, the Detector will enable the power
saving function automatically.
Configuring the OFF Delay
The Configuration parameter that can be used to adjust the amount of delay before
the OFF command is transmitted is Configuration Parameter # 2. This parameter
can be configured with the value of 1 through 127, where 1 means 1 second delay
and 127 means 127 seconds of delay.
Configuring the Phase Level of ON Command
The Configuration parameter that can be used to adjust the phase level of ON
command is transmitted is Configuration Parameter # 1. This parameter can be
configured with the value of 0 through 127.
Value 0: Set Device OFF(0x00)
Value 1-99: Set Device On (1-99)
Value 100-127: Set Device On to the last phase (0xFF)
Note:
0xFF means the device will be on to the last phase that the device was turned