background image

E2632

Rev 17.05.2021

If the gas concentration for each sensor exceeds the LOW alarm setpoint, the bottom red
LED of that sensor starts flashing at a rate of 1 Hz, and the relay RE1 switches over. The
first alarm stops automatically if the gas concentration drops below 80% of the LOW alarm
setpoint.

If the gas level exceeds the HIGH alarm setpoint, the bottom red LED of that sensor starts
flashing and the buzzer starts beeping at a rate of 2 Hz, and also the relay RE2 switches
over. Depending on the selected release mode, the HIGH alarm stops automatically or can
be stopped with a short touch of the magnet key, if the gas level has dropped below 80% of
the LOW alarm setpoint. Upon contact, the key should activate the reed switch located left
of the sensor inside the device.

Apart from the warming-up period, a 2....10 s touch causes the device to reset and perform
the self-diagnostic routine for testing purposes. To check the visual and acoustic alarm,
touch the device with the magnet key for more than 10 s. This will launch blinking and
beeping (stops as soon as the key is withdrawn).

Sensor probe handling

The sensor probes of all types are equipped with a hydrophobic microporous PTFE filter to
protect the sensor from dust, dirt, and water drops. The filter may be replaced if it gets
strongly contaminated. To replace the PTFE filter, unscrew the M25 nut and remove the old
filter. Place a new filter into the nut and tighten it again.

NOTE! Never stab or press the filter near its center where the sensor is located since this
may damage the sensor. Do not remove the filter as it may cause the device to show
incorrect values.

The recommended orientation of the sensor probes is vertical with the sensor tips pointing
downwards. This prevents the possible accumulation of condensed water on the sensor
protection filter.

The electrochemical gas sensor replacement procedures

1.

Un-power detector;

2.

Unscrew sensor head cap;

3.

Carefully take out the sensor cell;

4.

Disconnect the first wire from the old sensor and connect to the same pin on the
new sensor;

5.

Disconnect the second wire from the old sensor and connect to the second pin on
the new sensor;

6.

Insert the new sensor into the sensor head and screw back the sensor head cap.

10

Summary of Contents for E2632

Page 1: ...Two Channel Gas Detector E2632 User Manual...

Page 2: ...nts 5 Operating conditions 5 Installation guidelines 6 Mounting dimensions 7 Connections 8 Operation 9 Sensor probe handling 10 The electrochemical gas sensor replacement procedures 10 Maintenance 11...

Page 3: ...ioxide SO2 0 10 ppm 0 100 ppm Electrochemical 2 years 30 sec 6 months Ammonia NH3 0 100 ppm 0 500 ppm Electrochemical 2 years 50 sec 6 months Methane CH4 0 100 LEL Metal oxide semiconductor 5 years 60...

Page 4: ...max 5 A 30 VDC 250 VAC Alarm signaling Blinking LEDs buzzers 85 dB LEDs Green red operation fault red gas alarm Enclosure Grey ABS wall mount protection class IP65 Dimensions H123 W201 D55 mm Output c...

Page 5: ...components directly as they are sensitive to static electricity Connection diagrams can be found in the connections section The device might not perform correctly or be damaged if the wrong power sup...

Page 6: ...e area of the device However the coverage area for a detector is usually between 2 5 to 5 meters radius For early leakage detection install the sensor as close as possible to the potential leakage sou...

Page 7: ...E2632 Rev 17 05 2021 Mounting dimensions 7...

Page 8: ...of the external devices to each sensor Connect the power terminals N and L to the 24 V AC DC source if you are using detector version 24 or to 230 V AC mains if you are using detector version 230 see...

Page 9: ...the sensor to stabilize For stable operating it is recommended to keep the detector powered constantly except for periods of maintenance and calibration replacement etc Operation During the first ca 6...

Page 10: ...l launch blinking and beeping stops as soon as the key is withdrawn Sensor probe handling The sensor probes of all types are equipped with a hydrophobic microporous PTFE filter to protect the sensor f...

Page 11: ...fore delivery Provided that the sensor is used under moderate conditions field recalibration is recommended every few months The calibration intervals are stated in the gases table Please contact your...

Page 12: ...nufacturer will at its option either repair or replace a product that proves to be defective This warranty is void if the product has been operated in conditions outside ranges specified by the Manufa...

Reviews: