background image

RS485 communication interface

Databits: 8
Parity: none
Stop bits: 1 or 2

Supported Modbus functions:
03 - read multiple registers
06 - write single register

Modbus registers (0-based, decimal format)

Reg Description

Supported values

1 Hardware version

read only

2 Software version

read only

3 Product serial number

 read only

4 Slave ID (network address)*

1...247, default 1

5 Baudrate*

1200...57600 baud, default 9600

6 Response delay, ms

1...255 ms, default 10

7 Stop bits*

1 / 2, default 1

17  Restart

write ‘42330’ for soft restart

162 Temperature shift adjustment

-32000...+32000 T units (0,01 °C)

165 Gas channel shift adjustment

-32000...+32000 gas units, default 0

166 Gas channel slope adjustment 1...65535, default 512
167 Output change rate limit

1...32000 gas units/s, 0=no limit

168 Integrating filter time constant

1...32000 s, 0=no, integrating filter

201 Parameter assigned to OUT1

0=none, 1=T, 2=gas, 9=forced by 203

202 Parameter assigned to OUT2

0=none, 1=T, 2=gas, 9=forced by 204

203 Forced value for OUT1

0...1000 (0.0...100.0% of full scale)

204 Forced value for OUT2

0...1000 (0.0...100.0% of full scale)

211 Parameter assigned to RE1

0=none, 1=T, 2=gas, 9=forced by 213

212 Parameter assigned to RE2

0=none, 1=T, 2=gas, 9=forced by 214

213 Forced state for RE1

0=off, 1=on (relay control by Modbus)

214 Forced state for RE2

0=off, 1=on (relay control by Modbus)

215 Switch delay for RE1

0...1000 s, default 0

216 Switch delay for RE2

0...1000 s, default 0

217 Min on/off time for RE1

0...1000 s, default 0

218 Min on/off time for RE2

0...1000 s, default 0

219 Control logic for relay RE1

0:_, 1:_↓↑

P, 2:

P↑↓_, 3:_↑

P↓_,

4:

P↓_↑

P

220 Control logic for relay RE2

0:_, 1:_↓↑

P, 2:

P↑↓_, 3:_↑

P↓_,

4:

P↓_↑

P

221 LOW setpoint for relay RE1

-32000...+32000, gas or T units

222 HIGH setpoint for relay RE1

-32000...+32000, gas or T units

223 LOW setpoint for relay RE2

-32000...+32000, gas or T units

224 HIGH setpoint for relay RE2

-32000...+32000, gas or T units

258 Measured temperature

-4000...+12500 T units (0,01 °C)

259 Gas concentration

0...65535 gas units

261 0% value of OUT1

-32000...+32000 gas units / integer °C

262 100% value of OUT1

-32000...+32000 gas units / integer °C

263 0% value of OUT2

-32000...+32000 gas units / integer °C

264 100% value of OUT2

-32000...+32000 gas units / integer °C

* - the new value is applied after restart
Broadcast ID=0 may be used to assign a new ID to device with unknown ID

E2611-H2_UM_EN 

         Rev 05.03.2018

Specifications

Sampling method

Diffusion

Sensor type

Metal oxide
semiconductor

Pellistor (catalytic)

Detection range

Up to 100% LEL

Resolution

0,1% LEL

0,1% LEL

Response time

~ 60 seconds

~ 10 seconds

Sensor lifetime

> 5 years

> 2 years

Calibration interval

12 months

3...6 months

Operating
conditions

-30...+70 °C

-20...+60°C

<95 %RH, without condensation, non-ATEX
Normal ambient oxygen level/
Avoid strong mechanical shock, vibrations or EMI
Avoid exposure to corrosive gases or silicon containing
products

Warm-up time

up to 1 minute, depending on unpowered period and
atmosphere

Power supply

11...30 VDC/VAC
with integrated mains supply module 90...265 VAC

Power consumption

< 2 VA

Analog outputs

2 × 4-20 mA or 0-10 V

Load resistance

for 4-20 mA output mode: R

L

 < (Us - 2V) / 22 mA

for 0-10 V output mode: R

L

 > 250 kOhm

Digital interface

RS485, Modbus RTU protocol

Relay outputs

2 × SPST, max 5 A, 30 VDC / 250 VAC

Alarm

Buzzer 85 dBi

Enclosure

ABS plastic with ventilation slots, wall-mount,
protection class IP20

Dimensions

85 × 85 × 37 mm

Electromagnetic
compatibility

according to 2014/30/EU, 2014/35/EU and EN61321-1
standard requirements

Factory settings

Target gas

Hydrogen

Gas unit

0,1 % LEL

OUT1 assignment and scale

2: gas, 0 - 100% LEL

OUT2 assignment and scale

2: gas, 0 - 100% LEL

RE1 assignment and logic

2: gas, 1: on at high values

RE1 HIGH setpoint (set)

20% LEL

RE1 LOW setpoint (release)

16% LEL

RE2 assignment and logic

2: gas, 1: on at high values

RE2 HIGH setpoint (set)

50% LEL

RE2 LOW setpoint (release)

40% LEL

Evikon MCI OÜ  

Teaduspargi 7, Tartu 

info@evikon.eu

Tel +372 733 6310 

  50411 Estonia 

 www.evikon.eu

Hydrogen detector-transmitter

E2611-H2

User manual