NOTE: Disconnect line power before changing DIP switch settings. Failure to
disconnect power will damage electronics and void warranty.
All membrane sensors:
(FC1, TC1, CD7, OZ7)
In case the cell current is >200µA (e.g., in waters with high conductivity), change
the setting of positions 1 through 3 as follows:
DIP Switch (default 200µA):
DIP Switch
(only if sensor current >200µA to max. 1000 µA):
The other positions remain as before.
2.1.5 Installing the Depolox
®
3
plus
Electronics Module
The Depolox® 3
plus
electronics module is built in a wall-mounted housing and
should be mounted near the flow-through assembly.
2.1.6 Electrical Connections
Connect the sensor cables, alarm signal lines and the power cable according
to the wiring diagrams (see Dwgs. 50.560.155.010, .020 and .050).
Do not switch on the power.
WARNING: ONLy AUTHORIzED AND qUALIFIED ELECTRICIANS ARE PER-
MITTED TO INSTALL THE DEVICE AND OPEN THE HOUSING. THE UNIT
MAy ONLy BE PUT INTO OPERATION WHEN THE HOUSING IS CLOSED,
AND MUST BE CONNECTED TO EARTH GROUND. MODIFICATIONS TO
THE UNIT THAT GO BEyOND THOSE DESCRIBED IN THE MANUAL ARE
NOT PERMISSIBLE.
!
4 5 6 7 8 9 10
oFF
Summary of Contents for Wallace & Tiernan Depolox 3 plus
Page 11: ...WT 050 560 001 UA IM 0714 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Page 12: ......
Page 19: ......
Page 39: ...WT 050 560 001 UA IM 0714 26 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Page 40: ......
Page 109: ......
Page 127: ...WT 050 560 001 UA IM 0714 112 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Page 128: ......
Page 146: ...WT 050 560 001 UA IM 0714 130 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Page 147: ......
Page 157: ......