WT.050.560.001.UA.IM.0714
50
Depolox
®
3 plus ReSIDUAl ANAlYZeR
eVoQUA
W3T140188
• pH OFFSET Calibration
a. With sensor in flow block, test sample pH.
b. Press the
F
key until the “CALIBRATION” menu is displayed. Press the
key until the “pH offset” menu is found.
c. Press the
key.
d. Press
or
keys until the offset value is the difference of the displayed
pH value compared to the manually measured value. Store this value
using the
key. (Example: For pH7 solution, the basic display reads
7.15pH. The offset value will be adjusted to -0.15 because the value is
to come down from 7.15pH to 7.00pH.)
e. Return to the basic display by pressing the
eSC
key twice.
• pH Zero Calibration (Isothermal Intersection, pH7):
a. Starting at the basic display press the
F
key until the “CALIBRATION”
menu is displayed. Press the
key until the “pH zero” menu is found.
b. Withdraw the pH sensor. Fill 10 to 20 ml “pH7.00” buffer solution into
a 100 ml beaker. Dip the sensor at least 20mm deep in the buffer so-
lution pH7.00 and agitate lightly until the displayed pH value remains
constant.
c. Press the
key.
d. Press the
or
keys until the displayed value agrees with the value
of the pH buffer solution. Store this value using the
key.
• pH Span Calibration:
a. Rinse the sensor in distilled water (in order to avoid spreading the
buffer solution).
b. Press the key
until the “pH span” menu is reached.
c. Pour 10 to 20ml pH4 or pH10 buffer solution into another beaker. Dip
the pH sensor at least 20mm deep in the buffer solution and agitate
lightly until the displayed pH value remains constant. (In case another
buffer solution is used: the pH value of the buffer solution must be less
than pH6 or greater than pH8.)
d. Press the
key to open the “pH Span” menu.
Summary of Contents for Wallace & Tiernan Depolox 3 plus
Page 11: ...WT 050 560 001 UA IM 0714 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Page 12: ......
Page 19: ......
Page 39: ...WT 050 560 001 UA IM 0714 26 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Page 40: ......
Page 109: ......
Page 127: ...WT 050 560 001 UA IM 0714 112 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Page 128: ......
Page 146: ...WT 050 560 001 UA IM 0714 130 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Page 147: ......
Page 157: ......