d. Perform calibration.
3.3.3
Free Chlorine (Depolox
®
5 Bare Electrode) Flow Block Assemby
The Depolox® 5 bare electrode flow block assembly contains the following:
• Cell body with cover
• Flow control valve
• 3 electrode cell for Cl
2
, ClO
2
, O
3
, or KMnO
4
• Multi-sensor
• Drain
• Fine filter (For use only when membrane sensors are used.)
• Sample water inlet with check valve and ball valve
Two clips are installed in the housing cover. These are used to facilitate pH and
fluoride calibration. Insert these clips into the rear panel of the housing. Insert
one clip into the upper catch to attach the sensor. Insert the second clip into
the center catch inorder to measure the buffer or calibration solution using
the plastic cups provided.
The cell body can be equipped with up to five sensors.
A
B
C
D
E
F
G
H
I
J
K
A Cell body with cover
B Plastic housing
C Flow control valve
D 3 electrode cell for Cl
2
,
ClO
2
, O
3
or KMnO
4
E Drain
F Ball valve
G Fine filter
H Lower clip and cup
I
Multi-sensor
J
Upper clip
K Sensors
Summary of Contents for Wallace & Tiernan Depolox 3 plus
Page 11: ...WT 050 560 001 UA IM 0714 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Page 12: ......
Page 19: ......
Page 39: ...WT 050 560 001 UA IM 0714 26 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Page 40: ......
Page 109: ......
Page 127: ...WT 050 560 001 UA IM 0714 112 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Page 128: ......
Page 146: ...WT 050 560 001 UA IM 0714 130 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Page 147: ......
Page 157: ......