30° pitch on this side
B
Rounding on this side
Locking ring for PVC hose with 2 clamping points:
With rigid pipework:
1. Connect the sample water pipework to the ball valve connection threads.
2. Ensure that the sample water pipework is installed without mechanical
stress.
Connecting the sample water outlet:
NOTE: Never use copper tubing.
1. No back-pressure is permitted in the cell body. The sample water outlet
must be open.
2. The sample water outlet must be installed so that no siphon effect can
occur.
Recommendation: Position the outlet above the drain line opening.
With hose connection:
See hose connection instructions described in sample water inlet connection.
With rigid pipework:
1. Connect the sample water pipework to the connection nozzle.
2. Ensure that the sample water pipework is installed without mechanical
stress.
Connecting the drain:
1. Ensure that the drain screw is always closed.