VF / DSL Cable Qualifier
62
4. Press the RUN / STOP button to start the test. The results will be displayed on screen in an easy to
understand format. Users will be able to a plot of the effects of crosstalk.
To repeat the test, press the RUN/STOP button. You can also change option settings if you wish by
pressing the function button associated with TEST SETUP.
7.13 Isolation Test or Stress Leakage Test
Note: This is a Single-ended test.
The Isolation Test or Stress Leakage Test will allow user to determine the quality of the copper pairs
insulation (sheathing). Measures are done by injecting a 100 VDC signal onto the circuit under test and the
CableSHARK loops for break downs in the insulation. A good reading is >30 Mohm.
1. From the Main Menu, select CABLE TESTS (function button F1).
2. Choose ISOLATION TEST or STRESS LEAKAGE from the pop-up menu.
3. Using the SELECT NEXT or SELECT PREVIOUS function keys, select a Test Type of
Tip to Ring
4. Press the RUN / STOP button to start the test. A countdown timer will allow users how long the test has
left to finish and will indicate the current Isolation resistance. A failure will be indicated by a resistance
result of less than 30 Mohm.
To repeat the test, press the RUN/STOP button. You can also change option settings if you wish by
pressing the function button associated with TEST SETUP.
7.14 ADSL Auto Test
Note: This is a Single-ended test.
The ADSL Auto test quickly evaluates a cable to see if it is acceptable for ADSL based on pre-defined PASS
/ FAIL criteria. The ADSL auto test will perform a DMM test, Longitudinal Balance, Load Coil test, Capacitive
Imbalance, DMT test, and a TDR test.
1. From the Main Menu, select ADSL AUTO (function button F2). This test will take place automatically.
Summary of Contents for CableSHARK P3
Page 1: ......
Page 10: ...VF DSL Cable Qualifier viii...
Page 12: ...VF DSL Cable Qualifier x...
Page 26: ...VF DSL Cable Qualifier 14...
Page 36: ...VF DSL Cable Qualifier 24...
Page 64: ...VF DSL Cable Qualifier 52...
Page 80: ...VF DSL Cable Qualifier 68...
Page 184: ...VF DSL Cable Qualifier 172...
Page 192: ...VF DSL Cable Qualifier 180...
Page 224: ...VF DSL Cable Qualifier 212...
Page 228: ...VF DSL Cable Qualifier 216...
Page 262: ...VF DSL Cable Qualifier 250 CSA Loop 6 CSA Loop 7 CSA Loop 8 Extended CSA Loop 9...
Page 263: ...CableSHARK P3 User Guide 251 Extended CSA Loop 10 Mid CSA Loop 0 Mid CSA Loop 1 Mid CSA Loop 2...
Page 264: ...VF DSL Cable Qualifier 252 Mid CSA Loop 3 Mid CSA Loop 4 Mid CSA Loop 5 Mid CSA Loop 6...
Page 265: ...CableSHARK P3 User Guide 253 ANSI Loop 2 ANSI Loop 3 ANSI Loop 4 ANSI Loop 5...
Page 266: ...VF DSL Cable Qualifier 254 ANSI Loop 6 ANSI Loop 7 ANSI Loop 8 ANSI Loop 9...
Page 267: ...CableSHARK P3 User Guide 255 ANSI Loop 11 ANSI Loop 12 ANSI Loop 13 ANSI Loop 15...
Page 268: ...VF DSL Cable Qualifier 256...
Page 286: ...VF DSL Cable Qualifier 274...
Page 292: ...VF DSL Cable Qualifier 280...
Page 294: ...VF DSL Cable Qualifier 282...
Page 316: ...VF DSL Cable Qualifier 304...
Page 318: ...VF DSL Cable Qualifier 306...
Page 319: ...CableSHARK P3 User Guide 307 Notes...
Page 320: ...VF DSL Cable Qualifier 308...