VF / DSL Cable Qualifier
66
7.18 Resistance Fault Locator (RFL)
Resistance Fault Locator (RFL) testing is a powerful measurement that is able to locate shorts between tip to
ring, tip to ground or ring to ground. These faults can be called shorts, grounds or crosses. The RFL
measurement used by the CableSHARK is referred to as a “separate good pair” test. This means the user
must have a good pair to compare against.
1. Attached a test cable to the 2Wire TX/RX port of the CableSHARK and connect the Tip, Ring, and
Ground leads to the circuit. Attach the RFL test cable (Yellow and Green leads) to a separate good pair.
2. From the Main Menu, select CABLE TESTS (function button F1).
3. Choose RESISTANCE FAULT LOCATOR from the pop-up menu.
4. Press the RUN / STOP button to start the test. Follow the instructions on screen to determine where the
far end strap is to be placed.
5. Results will provide a user with a clear indication of the distance to the fault, the distance from the fault
to the strap and the distance from the CableSHARK to the strap.
To repeat the test, press the RUN/STOP button.
7.19 K-Test
K-test testing is a powerful measurement that is able to locate shorts between tip to ring, tip to ground or ring
to ground. These faults can be called shorts, grounds or crosses. This test does not need a separate good
pair like the RFL test does.
1. From the Main Menu, select CABLE TESTS (function button F1).
2. Choose K-TEST from the pop-up menu.
3. Press the RUN / STOP button to start the test. Follow the instructions on screen to determine how to
connect the test cable and where the far end strap is to be placed.
4. Results will provide a user with a clear indication of the distance to the fault, the distance from the fault
to the strap and the distance from the CableSHARK to the strap.
To repeat the test, press the RUN/STOP button.
Summary of Contents for CableSHARK P3
Page 1: ......
Page 10: ...VF DSL Cable Qualifier viii...
Page 12: ...VF DSL Cable Qualifier x...
Page 26: ...VF DSL Cable Qualifier 14...
Page 36: ...VF DSL Cable Qualifier 24...
Page 64: ...VF DSL Cable Qualifier 52...
Page 80: ...VF DSL Cable Qualifier 68...
Page 184: ...VF DSL Cable Qualifier 172...
Page 192: ...VF DSL Cable Qualifier 180...
Page 224: ...VF DSL Cable Qualifier 212...
Page 228: ...VF DSL Cable Qualifier 216...
Page 262: ...VF DSL Cable Qualifier 250 CSA Loop 6 CSA Loop 7 CSA Loop 8 Extended CSA Loop 9...
Page 263: ...CableSHARK P3 User Guide 251 Extended CSA Loop 10 Mid CSA Loop 0 Mid CSA Loop 1 Mid CSA Loop 2...
Page 264: ...VF DSL Cable Qualifier 252 Mid CSA Loop 3 Mid CSA Loop 4 Mid CSA Loop 5 Mid CSA Loop 6...
Page 265: ...CableSHARK P3 User Guide 253 ANSI Loop 2 ANSI Loop 3 ANSI Loop 4 ANSI Loop 5...
Page 266: ...VF DSL Cable Qualifier 254 ANSI Loop 6 ANSI Loop 7 ANSI Loop 8 ANSI Loop 9...
Page 267: ...CableSHARK P3 User Guide 255 ANSI Loop 11 ANSI Loop 12 ANSI Loop 13 ANSI Loop 15...
Page 268: ...VF DSL Cable Qualifier 256...
Page 286: ...VF DSL Cable Qualifier 274...
Page 292: ...VF DSL Cable Qualifier 280...
Page 294: ...VF DSL Cable Qualifier 282...
Page 316: ...VF DSL Cable Qualifier 304...
Page 318: ...VF DSL Cable Qualifier 306...
Page 319: ...CableSHARK P3 User Guide 307 Notes...
Page 320: ...VF DSL Cable Qualifier 308...