Introducing the CT440
6
CT440
Product Features
Scan span > 5 nm
The scan span is important for the absolute and relative accuracy. The largest span
leads to the most accurate results (the specifications are for a 100 nm span).
Below 5 nm, the precision cannot be guaranteed and a warning message is issued to
prevent such running condition.
Input power between 1 mW and 10 mW
The input power of the source is mostly important for the wavelength referencing. An
excessively low input power should be detected but if it affects too much the
wavelength referencing, the warning message on the input power will be replaced by
an error message on the referencing.
The CT440 checks most of these behaviors and issues a warning in non standard running
configurations.
Supported TLS
TUNICS Plus
TUNICS Purity
TUNICS Reference
TUNICS T100S
T100S-HP
TUNICS T100R
VIAVI SWS
Agilent