Setting Up Your CT440 for Measurements
CT440
29
Performing User Referencing
To perform power referencing:
1.
To obtain the optimum performance of the system, make sure the optical connectors
are perfectly clean (see
2.
Connect a verified optical jumper between the output port of the CT440 and the
detector port you want to reference.
3.
In the
Scan Parameters
window, define the appropriate scanning and laser settings
(see
4.
Manually Starting/Stopping a Scan
5.
On the graph, verify that the measurement has been performed.
6.
In the main window, click the
Referencing
button
The
Power Referencing
and
Analog In BNC
areas enable you to reference the wanted
detector.
7.
In the
Referencing
window, click the
Reference Detector
button corresponding to the
connected detector.
The detector is referenced, the
icon appears next to the referenced detector number. If
scan conditions are modified, you may be required to reference the system again.
I
MPORTANT
User power-referencing file is stored on the computer on which the CT440 is
connected. If you change computer you must perform the referencing again.