Setting Up Your CT440 for Measurements
CT440
31
Performing User Referencing
5.
In the
Referencing
window, in the
Wavelength Referencing
area:
5a.
In the
Measurement (nm)
field, enter the measured value.
5b.
In the
Reference (nm)
field, enter the expected value for this absorption line
(in nm).
6.
Click the
Apply
button.
At the next scan, the software will apply the appropriate offset in wavelength to correct
the wavelength calibration. The offset is not applied on existing traces.
Resetting Calibration
You can recover the initial factory calibration at any time. Calibration reset erases all
calibration modifications done on the wavelength referencing, on detector calibration and
on detector zeroing.
To recover the initial factory calibration:
In the
Referencing
window, click the
Reset Calibration
button.
All referencing operations performed on the CT440 are erased.