Performing Measurement Scans
46
CT440
Synchronizing the CT440 with External Measurements
To synchronize the CT440 with external measurements:
1.
Using the coaxial cable, connect the Trigger In port of the remote instrument receiving
the TTL signal to the
Trigger Out
BNC connector located on the rear panel of the CT440
(see
2.
In the
Scan Parameters
window, select
Single
scanning mode.
3.
Click
Scan
to run a single scan.
The CT440 generates a TTL periodic signal during the measurement acquisition, as
explained in the above
.