Analyzing Traces
CT440
67
Defining the Analysis Parameters
IN/OUT Band Distance
(only if
In/Out Band Method
is set to
Set Distance
)
Spectral spacing in nm/THz between the in-band reference point and the out-band
reference points to be used for the crosstalk calculation. Default value: 1 nm
Average Loss & Ripple Settings
Averaging Range
Spectral range used in the analysis of in-band and out-band average loss and ripple.
Fixed Range
: provides a fixed calculation span (see
Calculation Span
parameter).
% Bandwidth
: sets the range to a fraction of the bandwidth measured from the
Spectral Width 1
tool (see
Loss
Wavelength or Frequency
0
OUT Ref
Avg Loss OUT
OUT Ref
Avg Loss OUT
Slope
Ripple IN
Calculation Span
Avg Loss IN
Lo
ss
Wavelength or Frequency
0
IN Ref
Avg Loss IN
OUT Ref
Avg Loss OUT
OUT Ref
Avg Loss OUT
Slope
Ripple IN
Bandwidth 1
% of Bandwidth 1