Analyzing Traces
74
CT440
Defining the Analysis Parameters
Roll-Off & Transition Band Settings
Transition Reference
Reference point to be used in the transition calculation:
In-Band
(default): the transition band is defined as the part of the trace between
Level@
Transition Reference
-
Min Exclusion Thresh
. and Level@
Transition
Reference
-
Max Exclusion Thresh
.
Out-Band
: the transition band is defined as the part of the trace between Level@
Transition Reference
+
Min Exclusion Thresh
. and Level@
Transition Reference
+
Max Exclusion Thresh
.
Min Exclusion Thresh
.
(in dB) Minimum threshold for the exclusion of data outside of the transition band.
Default value: 3 dB
Max Exclusion Thresh
.
(in dB) Maximum threshold for the exclusion of data outside of the transition band.
Default value: 20 dB.
Wavelength or Frequency
0
Lo
ss
Avg Loss IN
Max Roll Off
Tr
ansition Band
WL
max
Roll Off @ XdB
Min Exclusion Thresh.
Max Exclusion
Thresh.
Wavelength or Frequency
0
Lo
ss
Avg Loss OUT
Max Roll Off
Tr
ans
ition B
and
WL
max
Roll Off @ XdB
Min
Exclusion
Thresh.
Max Exclusion
Thresh.