Troubleshooting
84
CT440
Warning & Error Messages
[warning code 102]
Scan speed too high
The mean scanning
speed of the TLS is too
high (> 120nm/s).
The reference file is not
located in the correct
folder.
Decrease the scan speed.
Make sure the reference file is located in
C:\Users\Public\Documents\EXFO\CT440\
Make sure the reference file is located in
C:\Users\Public\Documents\EXFO\CT440\
[warning code 103]
High dynamical changes
at low level, reduce scan
speed
The scan speed and input
power are not suited for the
measurement.
Decrease the scan speed.
[warning code 104]
Unexpected source
behavior, check TLS
performance
The internal wavelength
referencing has detected a
troubling behavior.
Make sure the source performances meet
the requirements detailed in
Make sure that the input power meet the
requirements detailed in
[warning code 106]
Low power on one TLS
input
The power in the input port
is below -16 dBm.
Make sure the jumper is clean and
properly connected to the optical input of
the CT440.
Make sure the TLS performances meet the
requirements detailed in
Error or warning
Possible Cause
Possible solution