TDR Test
194
FTB-610
Manual TDR
Trace Setup
The
Trace Setup
button opens a page that allows you to configure the
trace parameters. Trace 2 and Trace 3 can be individually
enabled/disabled. Trace 1 is a
live
TDR trace, and is always enabled.
Note:
If two traces are shown, they can either be Trace 1 and Trace 2 or Trace 1
and Trace 3, not Trace 2 and Trace 3 since Trace 1 is always enabled.
To setup the trace parameters:
1.
From the
Manual TDR
page, select
Trace Setup
to open the page.
2.
In the
Trace Setup
page, select the parameters as required.
3.
Tap
Close
to confirm and close the page.
Trace 1 (Live) Type
can be configured as follows:
Normal
displays a regular live TDR trace.
Differential (1-2)
is the subtraction of
Trace 2
from the normal
TDR signal. The differential trace is a live trace.
Differential (1-3)
is the subtraction of
Trace 3
from the normal
TDR signal. The differential trace is a live trace.