Signal Test
220
FTB-610
WB Balance Test
Setting Threshold Values
The
Test Thresholds
page allows you to set the
Marginal
/
Good Balance
thresholds
for ADSL/2+ and VDSL2-8/12/17/30 WB Balance.
To set the Threshold values:
1.
From the
Copper
main menu
Test Groups
, select
Signal
test
.
2.
From the
Tests
menu, select
WB Balance
test.
3.
Tap
WB Balance
tab for which you want to set the threshold values.
4.
Tap
Edit Thresholds
button to open the
Test Thresholds
page.
5.
Set the threshold values as required.
6.
Tap
Close
to confirm and exit the page.
The page allows you to set the following parameters:
Threshold
is the threshold limit for each pair:
Maximum
,
Minimum
, or
None
.
Marginal
/
Good Balance
are balance thresholds ranging from 5 dB to
99 dB..