Noise Tests
Copper Test
261
NEXT
Setting Spectrum Test Thresholds
The
Test Thresholds
page allows you to set the
Spectrum Mode
Thresholds
.
To set the Threshold values:
1.
From the
Copper
main menu
Test Groups
, select
Noise
test
.
2.
From the
Tests
menu, select
NEXT
test.
3.
Tap
Spectrum
tab.
4.
Tap
Edit Thresholds
button to open the
Test Thresholds
page.
5.
Set the threshold values as required.
6.
Tap
Close
to confirm and exit the page.
The page allows you to set the following parameters:
Threshold
is the threshold limit for each pair:
Maximum
,
Minimum
, or
None
which disables the threshold. In this case, the pass/fail status
indicator will not be displayed.
Crosstalk
allows you to edit the NEXT threshold values for all
bandwidths.