Technical Specifications
310
FTB-610
COPPER SPECIFICATIONS
a, b, c
(continued)
Single-ended frequency response (attenuation)
d
Distance range
100 m to 5000 m (300 ft to 16000 ft)
Frequency range
4.3 kHz to 30 MHz
Frequency uncertainty (accuracy)
±(50 ppm + 1 digit)
Level uncertainty (accuracy)
±2 dB typical for 2.2 MHz and 8 MHz ranges
±3 dB for VDSL2-12 and VDSL2-17
±4 dB for VDSL2-30 ranges
Resolution
0.1 dB
Horizontal scale
ADSL2+ = 2.208 MHz, VDSL2-8 = 8 MHz, VDSL2-12 = 12 MHz,
VDSL2-17 = 17.66 MHz, VDSL2-30 = 30 MHz
Vertical scale
0 dB to 100 dB
Resistive fault location (RFL)
Test type
Single pair (two wire) and separate good pair (four wire) and Küpfmüller (K-Test)
Fault detection
0 to 20 M
1
Resolution
Three digits
Loop resistance
10 k
1
maximum
Multiple cable sections
Five (includes gauge and temperature setting)
Fault location
Total resistance, near-end to fault resistance, fault to strap resistance (three
significant digits, least significant digit 0.1
1
).
Total length, distance to fault, distance from fault to strap
(three significant digits, least significant digit 1 m)
Uncertainty (accuracy)
±(0.1
1
+ 1 % x RTS)
K-Test uncertainty (accuracy)
± (1.0
1
+ 1 % RTS + (Rf1 + Rf2)/10 M
1
) - double fault
± (1.0
1
+ 1% RTS) - single fault
Notes
a. Subject to change without notice.
b. Typical, at 23 °C ± 3 °C, on batteries, with no USB connection.
c. Specifications based on 24 AWG (PE 0.5 mm) cabling.
d. Specification based on 1 kft 24 AWG cabling. Range depends on cable type and condition.
TECHNICAL SPECIFICATIONS
Display
Color touchscreen, 800 x 480 TFT, 178 mm (7 in)
Interfaces
Two USB 2.0 ports
RJ45 LAN 10/100/1000 Mbit/s
Fiber inspection probe connector port (video)
Built-in Bluetooth and Wi-Fi (hardware option)
Five-color coded 2 mm analog safety shrouded line interfaces
Storage
8 GB internal memory (flash)
16 GB internal memory (flash), optional
Batteries
Rechargeable lithium-ion batteries
Operating time: 4.75 h (typical with extended battery)
GENERAL SPECIFICATIONS (MODULE ONLY)
Size (H x W x D)
130 mm x 252 mm x 56 mm (5
1
/
8
in x 9
15
/
16
in x 2
3
/
16
in)
Weight
0.93 kg (2 lb)
Temperature
operating
storage
0 °C to 40 °C (32 °F to 104 °F)
a
–40 °C to 70 °C (–40 °F to 158 °F)
Note
a. DC voltage, isolation resistance, VF and WB receiver = 0
o
C to 45
o
C (32
o
F to 113
o
F).