Analyzing Traces with the Bidirectional Analysis Application
214
FTB-7000 Series
Viewing Results
Information specific to the A->B or B->A trace is also displayed:
³
Range
: Acquisition range.
³
Duration
: Duration of the acquisition.
³
Span ORL
: ORL calculated either between the span start and the span
end, or on the total fiber span, depending on the option you have
selected in the
Setup
window.
³
High resolution
: Indicates whether or not the acquisitions were
performed using the high-resolution feature.
³
IOR
: Index of refraction of the displayed trace.
³
Backscatter
: Rayleigh backscatter coefficient of the displayed trace.
³
Helix factor
: Helix factor setting of the displayed trace.
³
Splice loss detection
: Splice loss threshold for detecting small
non-reflective events during trace analysis.
³
Reflectance detection
: Reflectance threshold for detecting small
reflective events during trace analysis.
³
End-of-fiber detection
: End-of-fiber threshold for detecting important
event loss, which could compromise signal transmission, during trace
analysis.