Maintenance
274
FTB-7000 Series
Verifying Your OTDR
To measure the dynamic range:
1.
Connect the OTDR as indicated below. Other configurations are
possible, such as the one explained in the section on how to determine
measurement range, if you use the shortest fiber length from that
setup. In all cases, the fiber should have several sections longer than
2 km, with no loss greater than 8 dB and with an average attenuation
not exceeding 1 dB/km.
Ensure the OTDR port and connectors are correctly cleaned, and that
the fiber settings are accurate (IOR, Helix factor, and RBS).
1 to M reels of fiber
( Add as needed )
OTDR
2km or more
Fiber End
2km or more
2 km or more
2 km or more
Fiber end
1 to M reels of fiber
(Add as needed)