Testing Fibers in Advanced Mode
42
FTB-7000 Series
Setting the Autorange Acquisition Time
Setting the Autorange Acquisition Time
Note:
This function is available both in Advanced and Auto modes.
When performing automatic acquisitions in Advanced mode (see
on page 37) or before activating Auto mode (see
on page 33), you can set an autorange
acquisition time for the OTDR to average acquisitions over a set time
period.
The application uses this value to determine the best settings for the test.
Note:
In Template mode, the acquisition time of the reference trace is used for all
trace acquisitions, not the autorange acquisition time.
To set the autorange acquisition time:
1.
From the button bar, press
OTDR Setup
then go to the
Acquisition
tab.
2.
Go to the
Autorange acquisition time
box and press the up or down
arrow to select your preference. The default value is 30 seconds.
3.
Press
Exit OTDR Setup
to return to the OTDR application.