SyncE Test Setup and Results
NetBlazer
207
Summary (SyncE)
The global test pass/fail verdict is displayed next to the
Test Status
field
when enabled (see
Pass/Fail Verdict
on page 204). The verdict is Pass
when all verdicts are Pass, the verdict is Fail as soon as one of the
verdict is Fail. The global pass/fail verdict is based on the following
criteria:
ESMC Rate Threshold
(when enabled),
QL Mismatch
Monitoring
(when enabled),
ESMC Loss
, and
Link Down
.
Start Time
indicates the time the test was started.
ESMC
Indicates the current (meter and value), average, minimum, and
maximum ESMC frame rate (including information and event frames)
received in frame/s. A fail verdict is also displayed beside the minimum
and/or maximum values when the ESMC frame rate is outside the range
while the Pass/Fail Verdict is enabled.
ESMC RX Rate
meter displays the current ESMC frame rate (including
information and event frames) in frame/s averaged over the last 11 frames
received. The pass/fail verdict is also displayed when enabled.
Note:
When the
Pass/Fail Verdict
and
ESMC Rate Threshold
check boxes are
selected (see page 204), the green region is delimited from 0.8 to 10.2
frames/s corresponding to a
PASS
verdict. The red regions above and
beyond the threshold correspond to a
FAIL
verdict. There is neither green
nor red region when the Pass/Fail verdict is disabled.
Summary of Contents for FTB-860
Page 7: ...NetBlazer vii B Glossary 371 Acronym List 371 VLAN 379 1588 PTP 380 SyncE 385...
Page 12: ......
Page 16: ......
Page 20: ......
Page 30: ......
Page 84: ......
Page 110: ......
Page 122: ......
Page 152: ......
Page 166: ......
Page 182: ......
Page 192: ......
Page 224: ......
Page 366: ......
Page 376: ......
Page 382: ...Specifications 370 FTB 860 880 Straight Through Cable 1000 Mbit s Crossover Cable 1000 Mbit s...