Remote Control Commands
Variable Attenuator
93
Specific commands-Quick Reference Command Tree
ATTenuation?
Description
This query returns a value indicating the current attenuation
setting OR the minimum/maximum attenuation setting.
Syntax
ATT? [<wsp><value>]
Parameters
The <value> parameter is optional and can be
“MIN” to obtain the minimum setting for the current wavelength
“MAX” to obtain the maximum setting for the current
wavelength
Response
The attenuation value in the format “999.99”. The units are dB.
Note that, depending on the current measurement mode, the
response will include different parameters:
In Absolute mode:
In Reference mode:
In Offset mode:
Example
ATT? MIN
Note
The ATT? command is equivalent to the INP:ATT? command.
Both give the same result.
response
actual attenuator general offset (CAL?)
+
=
response
actual
attenuation
reference value
INP:OFFS?
(
)
general offset
(CAL?)
+
–
=
response
actual
attenuation
wavelength-specific
offset
general offset
+
+
=
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for FVA-3100
Page 11: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 17: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 21: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 61: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 81: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 85: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 123: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...