Remote Control Commands
Variable Attenuator
95
Specific commands-Quick Reference Command Tree
D
Description
This command controls the shutter, which can block optical
continuity.
Syntax
D<wsp><numeric>
Parameters
The <numeric> parameter can be
“0” to open the shutter
“1” to close the shutter
Example
D 1
Note
This command cannot be issued while a program is running.
The D command is similar to the OUTP:STAT command.
D?
Description
This query returns the current shutter state.
Syntax
D?
Response
The shutter state in the format
“0” if the shutter is optically open
“1” if the shutter is optically close
Example
D?
Note
The D? command is similar to the OUTP:STAT? command.