Remote Control Commands
102
FVA-3100
Specific commands-Quick Reference Command Tree
INPut:OFFSet?
Description
This query returns the current reference value for the current
wavelength.
Syntax
INP:OFFS?
Response
The current reference in the format “999.999”. The units are dB.
Example
INP:OFFS?
INPut:STEP
Description
This command is used to set the size of the step between two
attenuation settings. Use the INP:STEP:CATA? query to obtain
the list of available step sizes. If the requested step is not
available, the FVA-3100 is set to the next available step. If the
requested step is greater than the maximum available step, the
FVA-3100 is set to the maximum available step.
Syntax
INP:STEP<wsp><step>[<wsp>dB]
Parameters
The <step> parameter is the requested step size. The unit of
measurement (dB) is optional.
Example
INP:STEP 0.1
Note
This command cannot be issued while a program is running.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for FVA-3100
Page 11: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 17: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 21: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 61: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 81: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 85: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 123: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...