Cleaning Connectors Equipped with EUI/EUA Adapters
To clean EUI/EUA adapters:
1.
Remove the EUI/EUA adapter from the module connector.
2.
Moisten a cleaning tip (2.5 mm tip) provided by EXFO with
only one
drop
of isopropyl alcohol.
3.
Slowly insert the cleaning tip into the adapter until it comes out on the
other side (a slow clockwise rotating movement may help).
I
MPORTANT
Alcohol may leave traces if used abundantly. Avoid contact between
the tip of the bottle and the cleaning tip, and do not use bottles
that distribute too much alcohol at a time.
Push
Turn
Pull
1
2
3