Remote Control Commands
Variable Attenuator
83
IEEE 488.2 Required Commands
*ESR?
Description
This query reads the contents of the Standard Event Register.
Syntax
*ESR?
Response
Binary integer with a value between 0 and 255.
*IDN?
Description
This query reads the FVA-3100 identification string.
Syntax
*IDN?
Response
“EXFO E.O. Engineering FVA-3100 Vxx.xx”, where xx.xx is the
current product version.
Note
The SYST:VERS? and IDN? commands are equivalent to the
*IDN? command. They give the same result.
*OPC
Description
This command causes the FVA-3100 to generate the operation
complete message in the Standard Event Status Register (ESR)
when all pending selected FVA-3100 operations have been
finished.
Syntax
*OPC
Example
*OPC,*IDN?