Remote Control
40
IQ-2600B
Specific Commands
SOURce:SWEep:CENTer?
Description
This query returns the central wavelength for the current sweep
program in the current spectral units (nm or THz).
Syntax
SOUR:SWE:CENT?
Response
The central wavelength for the current sweep program in the
current spectral units in the “9999.999” (nm) or “999.9999”
(THz) format.
See also
SOUR:SWE:MODE, SOUR:SWE:MODE?, SOUR:SWE:COUN,
SOUR;SWE:COUN?, SOUR:SWE:REP, SOUR:SWE:REP?,
SOUR:SWE:REV and SOUR:SWE:REV?
SOURce:SWEep:COUNt
Description
This command is used to specify how many times you want the
sweep program to loop. To specify whether or not you want the
sweep program to loop, use the SOUR:SWE:REP command.
Syntax
SOUR:SWE:COUN<space><count>
Parameters
The <count> parameter represents the new number of loops
in the “99” format. For continuous repetition, enter “0”.
Example
SOUR:SWE:COUN 32
See also
SOUR:SWE:MODE, SOUR:SWE:MODE?, SOUR:SWE:CENT,
SOUR;SWE:COUN?, SOUR:SWE:REP, SOUR:SWE:REP?,
SOUR:SWE:REV and SOUR:SWE:REV?