Remote Control
62
IQ-2600B
GPIB Troubleshooting
Receive “Undefined
header” error.
(continued)
Incorrect module
address.
Ensure that the correct
module address is being
used.
Incorrect command
syntax.
Verify and correct syntax.
Unstable
communications.
DMA is enabled.
Try with DMA disabled.
Incorrect termination
character.
Synchronize termination
characters between the
GPIB controller and the
SCPI Manager.
Commands are
working fine but the
results and
measurements are
not as expected.
Synchronization
problem with
commands in program.
Use the *OPC? query
after each command.
SCPI Manager unable
to communicate with
modules.
The module application
is not open.
Open the module
application.
Problem
Probable Cause
Solution
Table 4-4.
Troubleshooting (Part 2 of 2)