Maintenance
Tunable Laser Source
65
User Calibration Feature
To adjust your module at a certain wavelength,
1.
Access the
Setup
window by pressing the
Sweep
button
(F7)
.
2.
Click on the
Maintenance
tab. The following window appears:
Figure 5-1.
User Calibration Window
3.
Connect your IQ-2600B to a calibrated wavelength meter and enter the
measured wavelength in the
Wavelength
field.
4.
Click on
Adjust
. Your unit will verify if there is a difference between the
value you have entered and its current wavelength position. If there is a
difference, it will add an offset to compensate.
The difference between the pre-selected wavelength and the measured
wavelength cannot be greater than ±0.200 nm. If the difference between
the measured and pre-selected wavelengths is greater than ±0.200 nm (for
example, if you have entered the wrong value), the following window will
appear: