Troubleshooting
218
IQS-600
Solving Common Problems
The controller unit does not
detect the IQS modules in
your expansion unit.
The expansion unit is
not turned on.
Turn on the expansion
unit.
The modules are not
properly inserted into
the expansion unit.
Ensure that the modules
are properly inserted into
the unit (remember to
turn off your unit first).
The IQS interface
cable is not properly
connected.
Ensure that the cable
linking your expansion
unit with the main unit is
properly connected.
Non-compatible
modules are inserted
into a high-speed
expansion unit
(bearing the “HS”
inscription on its front
panel).
Ensure that the modules
are compatible with
high-speed units. For a
complete list of
supported modules,
contact the technical
support team (see
Contacting the Technical
Support Group
The standard module
inserted into your high-speed
controller or expansion unit is
not detected.
The module is probably
not compatible with
high-speed units (bearing
the “HS” inscription on
their front panels).
Ensure that the module is
compatible with high-speed
units. For a complete list of
supported modules, contact
the technical support team
Problem
Probable Cause
Recommended Action
Summary of Contents for IQS-600 Series
Page 1: ...User Guide Integrated Qualification System IQS 600...
Page 22: ......
Page 28: ......
Page 106: ......
Page 132: ......
Page 194: ...Using IQS Products in an Automated Test Environment 186 IQS 600 Standard Status Data Structure...
Page 252: ......
Page 274: ......
Page 336: ......
Page 366: ...Using the IQS 9600 Utility Module 358 IQS 600...
Page 367: ...Using the IQS 9600 Utility Module Integrated Qualification System 359...
Page 368: ......