Traffic Analyzer Tabs
182
IQS-8500 Series and IQS-8120NGE/8130NGE
Traffic Filter Configuration
Traffic Filter Configuration
Allows gathering statistics according to the programmed filters. Up to 10
filters having up to four operands each can be defined and enabled.
Note:
The
Filter Cfg
tab is only available when the Advanced Traffic Filtering
software option (SK-ADV-FILTERS) is enabled. Refer to Software Options
Press
TEST
,
Traffic Analyzer
, and
Filters
Cfg
.
Filter Number (Filter No.)
Allows the selection of the filter number (
1
to
10
).A filter used (enabled)
for data capture will not be configurable and the message
Filter in use for
data capture
Filter Configuration
The filter configuration section allows to configure the filter criteria for the
selected filter. The configuration is only possible when the
Enable
check
box is not selected.
“
(
“ and “
)
“: The open and close parenthesis may be useful to control the
precedence of operands when more than two operands are used. Only
one level of parenthesis is supported. When no parenthesis are used, a
logical
AND
will have precedence over a logical
OR
.
Not
: When selected, add the logical negation (not equal) operator for the
operand filter defined at its right.